Home | Products | FAQs | Interfacing to Test Equipment | Common Product Features | Submit Feedback | Employment


 

Models LE-01, LE-05, LE-15 Laser Extensometers

These extensometers are high precision non-contacting units for strain measurement in materials testing. They use a high speed laser scanner to measure the spacing between reflective tape strips on the sample. The measurement range is from 0.3 to 3.2 inches (8 to 81 mm) on the LE-01, 0.3 to 5 inches (8 to 127 mm) on the LE-05, and 0.3 to 15 inches (8 to 381 mm) on the LE-15. The gage length is determined by the user. This allows high elongation measurements when shorter gage lengths are used. The high resolution also allows accurate measurements of smaller strains.

The self-contained extensometer uses state-of-the-art laser diode technology. A digital display is included. The analog output may be used to connect to existing test controls. The RS-232 serial communications interface provides two way communications. Operation may be local or remote.

The scanning beam is always perpendicular to the specimen, unlike most laser extensometers. This eliminates errors when viewing through windows in chambers. It also minimizes sensitivity to the distance between the extensometer and the sample. Because the unit measures reflected light, no receiver is required behind the sample.

Easy to use

The visible laser light is simply aimed at the specimen, which has small reflective tape strips set at the gage length desired. The extensometer displays the actual measured gage length. If desired, the zero button will off set the output to zero. As the specimen is tested, the display will then read the elongation directly.

The analog output and RS-232 interface are easily connected to existing controls or data acquisition systems.

High temperature clip-on reflectors may be used as an alternate to tape reflectors. The reflective tape can be used at temperatures up to 300°F (150°C).  These are re-useable and available as an option.  They are rated for use to 800°F (425°C).

Features:

  • Non-contacting design requires only reflective tape marks on the specimen, or clip-on reflectors
  • Ideal for use in chambers-calibration not affected by aiming through viewing windows.
  • High resolution of 1 micron.
  • Full 3.2 inch (81 mm), 5 inch (127 mm), or 15 inch (381 mm) measuring range allows high elongation measurements (e.g. 200% on a 1 inch gage length).

Specifications:

Measurement Range: LE-01: 0.3 to 3.2 inches (8 to 81 mm), LE-05: 0.3 to 5 inches (8 to 127 mm), LE-15: 0.3 to 15 inches (8 to 381 mm)
Resolution:

Display 100 microinches or 1 micron;

Analog output and RS-232 are 1 micron resolution on lower ranges;

LE-15: 0.001" (0.01 mm) resolution.

Linearity': LE-01: ±0.0002 inches (±0.005 mm), LE-05: ±0.0004 inches (±0.01 mm), LE-15: ±0.002 inches (±0.04 mm)
Repeatability': LE-01: ±0.0001 inches (±0.001 mm), LE-05: ±0.0002 inches (±0.005 mm), LE-15: ±0.002 inches (±0.04 mm)
Scan Rate: 100 scans/second
Target Distance: 10 inches nominal (254 mm), 20 inches max (508 mm)
Zero Suppression: Reading may be set to zero anywhere in measuring range
Scan Line Orientation: Vertical
Scan Averaging: Moving window averaging over a selectable number of scans
Analog Output: 16 bit, ±10 VDC standard
Full scale ranges: 0.1, 0.2, 0.5, 1.0, 2.0, 5.0, 10.0, or 20.0 inches, or 2.0, 5.0,10, 20, 50, 100, 200, or 500 mm (model dependent)
Digital Communications: RS-232 serial communications, standard 3 wire
Selectable baud rate: 9600, 4800, 2400, or 1200
Display: 2 line, 16 character digital display, backlit LCD
Power Input: 115 VAC ±10%, 50/60 Hz standard, 230 VAC ±10%, 50/60 Hz optional
Size: LE-01,05: 13 L x 7.1 H x 4.5 W inches (330 x 180 x 114 mm), LE-15: 27 L x 18 H x 7.7 W inches (668 x 457 x 196 mm)
Weight: LE-01,05: 10 lbs (4.5 kg ), LE-15: 45 lbs (20.4 kg)
Mounting Provisions: 1/4-20 UNC tapped holes in base (4)
Laser Source: Diode laser, 670 nM, <1 mW maximum scanned output
CDRH Certified Class II laser instrument
Options: Consult factory for options like reflective clips for high temperature use.

1Over optimum displacement range at calibrated distance

Note: Specifications measured at 25 °C and 50 percent rel. humidity.

Contact Information

For price, delivery or other information, please click here to submit our information request form.

Telephone: 307-733-8360

FAX 307-733-8375

Address 3975 S. Highway 89, Jackson, WY  83001 USA

e-mail:

General Information: info@epsilontech.com
Sales: sales@epsilontech.com
Customer Support and Service:
service@epsilontech.com

france flag.gif (2350 bytes) France    italia flag.gif (2350 bytes) Italia    India   中国
 
©2004, Epsilon Technology Corp. All rights reserved.

visa103x65.gif (806 bytes)      mc103.gif (2101 bytes)