DSM-Plus, Digital Strain Meter

A high accuracy, higher speed meter that allows multiple calibrated ranges. Optionally, the DSM-Plus can be configured with two channels to read force and strain.This is a simple, low cost way to obtain the data for stress-strain plots. The automatic recognition feature allows multiple extensometers to be calibrated individually on one meter.

The DSM-Plus is ideal for customers who own several extensometers. When a configured extensometer is plugged in, the meter automatically recognizes it. It is also possible to calibrate the same extensometer in multiple ranges. For example, it is frequently convenient to calibrate an extensometer to a range of 10% of the full scale measuring range to allow greater sensitivity for tests at small strains.

The dual channel version is very useful for tests which require two strain readings, such as measurement of Poisson’s ratio or r-value tests on sheet metal samples. It also is a low cost solution to upgrading older test systems. One channel is used for strain and the other for the force reading from a load cell. The RS232 digital output or the analog outputs make computerized data acquisition simple.

A PDF document containing specifications on the DSM-Plus can be downloaded from the following link: PDF DOWNLOAD.

For price, delivery or other information, please click here to submit our information request form.

Features:

  • Automatic recognition of extensometers.
  • Accuracy: 0.01% of full scale ±1 digital count
  • 6 digit display in engineering units such as percent strain or
  • displacement.
  • Analog output with capability of 0 to ±10 VDC output
  • Common mode rejection: 115dB
  • Operating temperature range: 0 to 50 °C
  • Front panel tare button to zero output at the start of every test.
  • 60 readings per second update rate
  • Includes power cord, connector for extensometer(s), output cable and calibration with extensometer.
  • Input power: 110 VAC, 60 Hz, Optional 220 VAC, 50 Hz.
  • Optional second channel allows for second extensometer to be used simultaneously or load and strain to be measured simultaneously for a low cost solution to obtaining stress-strain plots for older machines.

DSM-Plus Available Options:

  • second channel for strain or load cell

Contact Information

For price, delivery or other information, please click here to submit our information request form.

Telephone: 307-733-8360

FAX 307-733-8375

Address 3975 S. Highway 89, Jackson, WY  83001 USA

e-mail:

General Information: info@epsilontech.com
Sales: sales@epsilontech.com
Customer Support and Service:
service@epsilontech.com

france flag.gif (2350 bytes) France    italia flag.gif (2350 bytes) Italia    India   中国
 
©2004, Epsilon Technology Corp. All rights reserved.

visa103x65.gif (806 bytes)      mc103.gif (2101 bytes)